PROVA 133

Documenting Multifunction Calibrator

Availability: In stock
  • Source 4~20 mA loop Current.
  • Simulate 4~20 mA transmitter.
  • Simulate electronic load (Max. 30V, 20mA).
  • Test LED brightness (0~24mA).
  • Source 0~70mV and 0~15V.
  • Calibrate temperature with selection of 11 types of thermocouples.
  • Source frequency (0.3 to 20KHz) of many waveforms.
  • Source: mA, V, Hz, sine wave, square wave, triangular wave, truncated sine wave, user programmable waveform and temperature.
  • Generate arbitrary waveform.
  • Generate single pulse (3μS to 999.99mS).
  • Map 4~20 mA into engineering units.
  • Measure: Current (mA), Voltage (mV, V), and temperature (°C, °F).
  • Measure 4~20 mA with 24V loop supply simultaneously.
  • Selectable 250Ω HART resistor to facilitate use with HART communication device.
  • Programmable cold junction compensation for temperature measurement.
  • Auto step and auto ramp for sourcing mA, V, and temperature.
  • Dot matrix LCD with backlight.
  • Rechargeable lithium battery (1600mAh).
  • Data logging function for source and measurement.
  • Program calibrator through PC USB port.
  • Programmable 0% and 100% value for easy 25% step function.
  • DTMF (Dual Tone Multi-Frequency) can perform professional testing for telephone line and audio product (MP3 or MD).
    • Shipping
    • Payments
    • Quantity
    • US$ 952Log in for sale price
    To show price, please join membership and log in.
    Documenting Multifunction Calibrator
    Documenting Multifunction Calibrator
    • Specification
    • Application

     

    Electrical Specifications:
    (23+/- 5, 10 minutes after turning on the power)

     

    mA (source)  (Vopen > 24V)

    Range

    Resolution

    Accuracy of Reading

    0.005mA to 4mA

     

    1uA

     

    +/-0.03% +/-5dgts

    4mA to 20mA

    +/-0.03% +/-3dgts

    20mA to 24mA

    +/-0.03% +/-5dgts

     

    V (source)  (maximum load 1mA, short circuit protection < 100mA)

    Range

    Resolution

    Accuracy of Reading

    0.005V to 10V

    0.001V

    +/-0.03% +/-5dgts

    10V to 15V

     

    mA (measure)

    Range

    Resolution

    Accuracy of Reading

    -4mA to -0.005mA

     

    1uA

     

    +/-0.03% +/-10dgts

    0.005mA to 4mA

    +/-0.03% +/-5dgts

    4mA to 20mA

    +/-0.03% +/-3dgts

    20mA to 24mA

    +/-0.03% +/-5dgts

    If reading of mA (measure) is less than 5 digits, it is displayed as 0.

     

    V (measure)

    Range

    Resolution

    Accuracy of Reading

    -3V to -0.005V

     

    0.001V

     

    +/-0.03% +/-10dgts

    0.005V to 10V

    +/-0.03% +/-5dgts

    10V to 24V

    +/-0.03% +/-5dgts

    If reading of V (measure) is less than 5 digits, it is displayed as 0.

     

    Frequency (source, 10 Vpp, 0V offset, square wave, duty cycle = 50%)

    Range (Hz)

    Input Resolution

    Accuracy

    0.3 to 99.999

    0.1Hz

    0.002Hz

    10.00 to 999.99

    0.1Hz

    0.02Hz

    1000.0 to 9999.9

    0.1Hz

    0.2Hz

    10000 to 20000

    1Hz

    2Hz

     

    Voltage Peak to Peak for Sine Wave
    (Vpp, 0.3~20KHz, 50% duty cycle, sine wave, 0V offset)

    Range (V)

    Resolution

    Accuracy of Reading

    0.1 to 20V

    0.001V

    5% +/- 0.3V

     

    Voltage Peak to Peak for Non-Sine Wave  (Vpp, 0.3~20KHz, 0V offset)

    Range (V)

    Resolution

    Accuracy of Reading

    0.1 to 20V

    0.001V

    6% +/- 0.4V

     

    Voltage Peak to Peak (Vpp, 0.3~20KHz, 50% duty cycle, square wave, 0V offset)

    Range (V)

    Resolution

    Accuracy of Reading

    1 to 20V

    0.001V

    6% +/- 0.4V

     

    Voltage of Offset (Maximum Vpp < 20V)

    Range

    Resolution

    Accuracy of Reading

    -5V to 5V

    0.001V

    5% +/-0.5V +/-5%xVpp

     

    Temperature, Thermocouples
    (source and measure, 0.1C & 0.1F Resolution, Internal Cold Junction Compensation, thermocouple accuracy not included, 3 minutes after plugging in thermocouples.)

     

    °C

    °F

    Range

    Accuracy

    Range

    Accuracy

    K

    -200 to -150

    2.0

    -382 to -238

    3.6

    -150 to 0

    1.2

    -238 to 32

    2.1

    0 to 1000

    0.8

    32 to 1832

    1.4

    1000 to 1370

    1.2

    1832 to 2498

    2.1

    J

    -200 to -150

    2.0

    -382 to -238

    3.6

    -150 to 0

    1.0

    -238 to 32

    1.8

    0 to 1050

    0.7

    32 to 1922

    1.2

    E

    -200 to -150

    1.5

    -382 to -238

    2.7

    -150 to 0

    0.9

    -238 to 32

    1.6

    0 to 850

    0.7

    32 to 1562

    1.2

    T

    -200 to -150

    1.5

    -382 to -238

    2.7

    -150 to 0

    1.2

    -238 to 32

    2.1

    0 to 400

    0.8

    32 to 752

    1.4

    R

    0 to 500

    1.8

    32 to 932

    3.2

    500 to 1760

    1.5

    932 to 3200

    2.7

    S

    0 to 500

    1.8

    32 to 932

    3.2

    500 to 1760

    1.5

    932 to 3200

    2.7

    N

    -200 to 0

    1.5

    -328 to 32

    2.7

    0 to 1300

    0.9

    32 to 2372

    1.6

    L

    -200 to 0

    0.9

    -328 to 32

    1.6

    0 to 900

    0.7

    32 to 1652

    1.2

    U

    -200 to 0

    1.1

    -328 to 32

    1.9

    0 to 600

    0.7

    32 to 1112

    1.2

    B

    600 to 800

    2.2

    1112 to 1472

    3.9

    800 to 1000

    1.8

    1472 to 1832

    3.2

    1000 to 1820

    1.4

    1832 to 3308

    2.5

    C

    0 to 1800

    1.0

    32 to 3272

    1.8

    1800 to 2310

    1.5

    3272 to 4190

    2.7

    mV

    -10mV to 70mV

    0.05mV

    -10mV to 70mV

    0.05mV

     

    Duty Cycle (%, square wave, 10 Vpp, 0.3~20KHz)

    Range

    Resolution

    Rise Time of Vpp

    Fall Time of Vpp

    0 to 100%

    1%

    10μS max,

    5μS typical

    15μS max,

    7.5μS typical

    Pulse (square wave, 10 Vpp, Offset -5V~+5V)

    Range

    Resolution

    Rise Time of Vpp

    Fall Time of Vpp

    3.0μS to 9999.9μS

    0.1μS

    10μS max,

    5μS typical

    15μS max,

    7.5μS typical

    10.000mS to 99.999mS

    0.001mS

    100.00mS to 999.99mS

    0.01mS

    DTMF (Hz)

    Range (Hz)

    Resolution

    Accuracy of Reading

    0.3 to 99.999

    0.1Hz

    0.002Hz

    10.00 to 999.99

    0.1Hz

    0.02Hz

    1000.0 to 9999.9

    0.1Hz

    0.2Hz

    10000 to 20000

    1Hz

    2Hz

    DTMF  (%)

    Range (%)

    Resolution

    Accuracy of Reading

    0% ~ 100%

    1%

    5%

    DTMF  (Phase Angle)

    Range (°)

    Resolution

    Accuracy of Reading

    0 ~ 360

    1°

    100μS+1°

    DTMF  (Vpp, F1=F2,

    Range

    Resolution

    Accuracy of Reading

    5V ~ 20V

    0.001V

    10% +/-0.6V

    DTMF  (Offset, F1=F2,

    Range

    Resolution

    Accuracy of Reading

    -5V ~ 5V

    0.001V

    10% +/-0.6V +/-5%xVpp

     

    General Specifications:

    AC Power Adaptor:

    AC 110V or AC 220V, 50/60Hz input.

    DC 15V / 0.5A output

    Dimension:

    214.0 (L) x 98.7 (W) x 56.0 (H) mm

    8.4" (L) x 3.9" (W) x 2.2" (H)

    Weight:

    650g / 22.9oz (Batteries included)

    Operation Environment:

    0℃ ~ 50℃, 85% RH

    Storage Environment:

    -20℃ ~ 60℃, 75% RH

    Accessories:

    Carrying case x 1,

    User manual x 1,

    AC power adaptor x 1,

    USB cable x 1,

    Software CD x 1,

    Software manual x 1,

    K-type thermocouple (dual plugs) x 1,

    Alligator clips x 2 (black and red),

    Test leads x 2 (black and red),

    Rechargeable lithium battery (11.1V/ 1600mAh) x 1

     

     

    Related Products