Hold + Average
Max + Min
ΔMax + ΔMin
Peak Max + Peak Min
Δ Peak Max + Δ Peak Min
Electrical Specifications: (23℃±5℃)
DC Current
Range |
Resolution |
Accuracy |
Overload Protection |
60 A |
10mA |
±1.5%±2dgts |
DC 400A
|
0-150 A |
100mA |
±1.5%±2dgts |
|
150-300 A |
100mA |
±2.2%±2dgts |
|
300-400 A |
100mA |
±4%±2dgts |
DCA Peak MAX ±400A Accuracy + (±2%±50dgts)
AC Current (A1: True RMS, A3: Average Sensing)
Range |
Resolution |
Accuracy |
Overload Protection |
|
50/60 Hz |
40 - 500Hz |
|||
60A |
10mA |
±1.5%±3dgts |
±2.0%±4dgts |
AC 400A
|
0-150A |
100mA |
±1.5%±3dgts |
±2.0%±4dgts |
|
150-300A |
100mA |
±2.2%±3dgts |
±2.5%±4dgts |
|
300-400A |
100mA |
4%±3dgts |
5%±4dgts |
ACA Peak MAX ±400A Accuracy + (2%±50dgts)
LPF @1KHz -3db
DC Voltage (Overload Protection 600V)
Range |
Resolution |
Accuracy |
Input Impedance |
60V |
10 mV |
±1.5%±3dgts |
10MΩ |
600V |
100 mV |
DCV Peak MAX ±600V Accuracy + (±2%±50dgts)
AC Voltage (A1: True RMS, A3: Average Sensing) (Input Impedance 10 MΩ)
Range |
Resolution |
Accuracy |
Overload Protection |
|
50/60 Hz |
40 - 1KHz |
|||
60V |
10mV |
±1.5%±5dgts |
±2.0%±5dgts |
AC 600V |
600V |
100mV |
ACV Peak MAX ±850V Accuracy + (±2%±50dgts)
LPF @1KHz -3db
Continuity (Open voltage 0.4V, Overload protection AC 600V)
Range |
Resolution |
Accuracy |
Beeping |
60-600Ω |
0.1Ω |
±1.0%±2dgts |
<30.0Ω (approx.) |
Resistance (Ω) (Open voltage 0.4V)
Range |
Resolution |
Accuracy |
Overload Protection |
600Ω |
0.1Ω |
±1.5%±2dgts
|
AC 600V
|
6KΩ |
1Ω |
||
60KΩ |
10Ω |
||
600KΩ |
100Ω |
Frequency (auto range)
Range (Hz) |
Resolution (Hz) |
Accuracy |
Sensi- tivity |
Overload Protection |
4-99.99 |
0.01 |
±0.5%±0.02 |
V:3V/10V A:1A/5A |
AC 600V
AC400A |
100.0~999.9 |
0.1 |
±0.5%±0.2 |
V:2V/10V A:1A/5A |
|
1.000K-9.999K |
10 |
±0.5%±20 |
V:2V/10V A:1A/5A |
|
10.00K-99.99K |
1.2K |
±0.5%±2.4K |
V:2V/10V A:1A/5A |
|
100.0K-200.0K |
5K |
±0.5%±10K |
V:2V/10V A:1A/5A |
Temperature (for using K-Type thermocouples, ℃ or ℉)
Range |
Resolution |
Accuracy |
Overload Protection |
-40 – 0 ℃ |
0.1 ℃ |
±2.0%±2 ℃ |
AC 300V
|
0 – 400 ℃ |
0.1 ℃ |
±0.5%±2 ℃ |
|
-40 – 32 ℉ |
0.1 ℉ |
±2%±3.6 ℉ |
|
32 – 752 ℉ |
0.1 ℉ |
±0.5%±3.6℉ |
Diode Test
Range |
Resolution |
Accuracy |
Overload Protection |
0 – 1.999V |
0.001V |
±2.5%±5dgts |
AC 600V |
Capacitance
Range |
Resolution |
Accuracy |
Overload Protection |
600nF |
0.1nF |
±2.0%±0.2 nF |
AC 600V
|
6μF |
1nF |
±2.0%±0.02μF |
|
60μF |
10nF |
±2.0%±0.02μF |
|
600μF |
100nF |
±2.0%±0.2μF |
|
6000μF |
1μF |
±2.5%±5μF |
General Specifications: Indoor Use
Conductor Size: 25mm max. (approx.)
Battery Type: two 1.5V SUM-3
Display: 6000 counts
Range Selection: Auto and Manual
Overload Indication: OL
Power Consumption: 17 mA (approx.)
Low battery Indication:
Sampling Time: 2 times/sec.
Auto-power-off: 30 minutes
Operating Temperature: 0C to 50C
Operating Humidity: less than 85% relative
Altitude: up to 2000M
Storage Temperature: -20C to 60C
Storage Humidity: less than 75% relative
Dimension: 190mm (L) x 66mm (W) x 36mm (H)
7.5" (L) x 2.6" (W) x 1.45" (H)
Weight: 250g (battery included)
Accessories: Carrying bag x 1
Users manual x 1
1.5V battery x 2
Test leads x 1
K-type thermocouples x 1
Adapter (for Capacitance and K-type thermocouples) x 1
1. (1) Troubleshooting in
i. Automotive Maintenance
ii. electric vehicle
iii. plant maintenance
iv. HVAC
v. telecom works
vi. marine
vii. motorcycle repairs
(2) DC charging current of batteries of electric vehicles.
(3) DC current consumption of DC motors of Drones.
(4) DC current output of solar panels.
(5) Measurements of voltage, current, and frequency of inverters (VFD).
(6) Temperature measurements of abnormal components.