ProcessDMM (DMM + Multifunction Calibrator)
- Specification
- Application
Electrical Specifications:
(23+/- 5℃, 3 minutes after turning on the power)
DCmA SOURCE (Max. load 1KΩ, Max. voltage supply 24V)
Range |
Resolution |
Accuracy |
4 - 20mA, 0 - 20mA, 0 - 24 mA |
1μA |
± 0.025% ±3μA |
When output open, the LCD displays “OL”.
DCmV, DCV SOURCE (Max. load 1mA, Short circuit protection<100mA)
Range |
Resolution |
Accuracy |
0 - 100mV |
0.1mV |
± 0.05% ± 0.3mV |
0.1 - 1.0000V |
100μV |
± 0.05% ± 300μV |
1 - 12.000V |
1mV |
± 0.05% ± 3mV |
When output short circuit, the LCD displays “OL”.
ACmA MEASURE (Ture RMS)
Range |
Resolution |
Accuracy |
|
50/60Hz |
40-1KHz |
||
24mA |
1uA |
±1%±5dgts |
±2%±5dgts |
DCmA MEASURE
Range |
Resolution |
Accuracy |
24mA |
1uA |
±0.05%±2dgts |
DCV MEASURE (Overload protection AC 600V, Input impedance 10MΩ)
Range |
Resolution |
Accuracy |
2.4V |
0.1 mV |
±1%±3dgts |
24V |
1 mV |
±1%±3dgts |
240V |
10 mV |
±1%±3dgts |
600V |
100mV |
±1.5%±3dgts |
ACV MEASURE (Ture RMS)
(Overload protection AC 600V, Input impedance 10MΩ)
Range |
Resolution |
Accuracy |
|
50/60Hz |
40-1KHz |
||
24V |
1 mV |
±1%±5dgts |
±2%±5dgts |
240V |
10 mV |
±1%±5dgts |
±2%±5dgts |
600V |
100mV |
±1.5%±5dgts |
±2%±5dgts |
Frequency SOURCE (TTL, Square wave, Duty cycle = 50%)
Range (Hz) |
Resolution |
Accuracy |
1.0~9999.9Hz |
0.1Hz |
0.1Hz |
10000~20000Hz |
1Hz |
2Hz |
Frequency MEASURE (Sensitivity RMS 1V)
Range (Hz) |
Resolution |
Accuracy |
1.000~39.999Hz |
0.001Hz |
±0.5%±0.003Hz |
40.00~399.99Hz |
0.01Hz |
±0.5%±0.03Hz |
0.4000K~3.999KHz |
1Hz |
±0.5%±3Hz |
4.000K~39.999KHz |
1Hz |
±0.5%±3Hz |
40.00K~399.99KHz |
10Hz |
±0.5%±30Hz |
0.4000M~3.999MHz |
1KHz |
±0.5%±3KHz |
4.000M~39.999MHz |
1KHz |
±0.5%±3KHz |
Duty Cycle SOURCE/MEASURE (1~20KHz)
Range |
Resolution |
Accuracy |
1% to 99% |
1% |
±1%±30μS |
Resistance (Ω) MEASURE (Open voltage 0.4V, Overload protection AC 600V)
Range |
Resolution |
Accuracy |
400Ω |
0.01Ω |
±1%±2dgts |
4KΩ |
0.1Ω |
±1%±2dgts |
40KΩ |
1Ω |
±1%±2dgts |
400KΩ |
10Ω |
±1%±2dgts |
4MΩ |
100Ω |
±1.5%±2dgts |
40MΩ |
1KΩ |
±1.5%±2dgts |
Continuity MEASURE:
Open voltage 2V, Overload protection AC 600V, <10Ω Beep.
Diode MEASURE (Open voltage 2V, Overload protection AC 600V)
Range |
Resolution |
Accuracy |
0~1.9999V |
0.0001V |
±2.5%±5dgts |
Temperature Thermocouples
(SOURCE and MEASURE, 0.1℃ & 0.1℉ Resolution, Internal Cold Junction Compensation, thermocouples accuracy not included, 3 minutes after plugging in thermocouples.)
|
°C |
°F |
||
Range |
Accuracy |
Range |
Accuracy |
|
K |
-200 to -150 |
2.0 |
-382 to -238 |
3.6 |
-150 to 0 |
1.2 |
-238 to 32 |
2.1 |
|
0 to 1000 |
0.8 |
32 to 1832 |
1.4 |
|
1000 to 1370 |
1.2 |
1832 to 2498 |
2.1 |
|
J |
-200 to -150 |
2.0 |
-382 to -238 |
3.6 |
-150 to 0 |
1.0 |
-238 to 32 |
1.8 |
|
0 to 1050 |
0.7 |
32 to 1922 |
1.2 |
|
E |
-200 to -150 |
1.5 |
-382 to -238 |
2.7 |
-150 to 0 |
0.9 |
-238 to 32 |
1.6 |
|
0 to 850 |
0.7 |
32 to 1562 |
1.2 |
|
T |
-200 to -150 |
1.5 |
-382 to -238 |
2.7 |
-150 to 0 |
1.2 |
-238 to 32 |
2.1 |
|
0 to 400 |
0.8 |
32 to 752 |
1.4 |
|
R |
0 to 500 |
1.8 |
32 to 932 |
3.2 |
500 to 1760 |
1.5 |
932 to 3200 |
2.7 |
|
S |
0 to 500 |
1.8 |
32 to 932 |
3.2 |
500 to 1760 |
1.5 |
932 to 3200 |
2.7 |
|
N |
-200 to 0 |
1.5 |
-328 to 32 |
2.7 |
0 to 1300 |
0.9 |
32 to 2372 |
1.6 |
|
L |
-200 to 0 |
0.9 |
-328 to 32 |
1.6 |
0 to 900 |
0.7 |
32 to 1652 |
1.2 |
|
U |
-200 to 0 |
1.1 |
-328 to 32 |
1.9 |
0 to 600 |
0.7 |
32 to 1112 |
1.2 |
|
B |
600 to 800 |
2.2 |
1112 to 1472 |
3.9 |
800 to 1000 |
1.8 |
1472 to 1832 |
3.2 |
|
1000 to 1820 |
1.4 |
1832 to 3308 |
2.5 |
|
C |
0 to 1800 |
1.0 |
32 to 3272 |
1.8 |
1800 to 2310 |
1.5 |
3272 to 4190 |
2.7 |
Dimension |
214.0 (L) x 98.7(W) x 56.0(H) mm 8.4" (L) x 3.9" (W) x 2.2" (H) |
Weight |
650g / 22.9oz (batteries included) |
Operation Environment |
0℃ ~ 50℃ 85% RH |
Storage Environment |
-20℃ ~ 60℃ 75% RH |
Accessories |
Carrying case x 1 User manual x 1 1.5V SUM-3 Battery x 5 K-type thermocouples (dual plugs) x 1 Alligator clips x 2 (black and red) Test leads x 2 (black and red) |
1. Calibration of 4~20mA panel meters.
2. LED brightness testing (0~24mA).
3. Temperature calibration of panel meters or instruments for 11 types of thermocouples.
4. Calibration of valve opening by changing duty cycle of a PWM signal.
5. Output Frequency measurement of Inverter.