AC/DC mA Current Probe
- Specification
- Application
Electrical Specifications: (23℃±5℃)
AC / DC Current:
Range |
Output |
Accuracy |
|
DC 400mA1 |
DC 1mV/mA |
----------------1 |
|
DC 4A |
DC 100mV/A |
±2.0%±3mA |
|
DC 30A |
DC 10mV/A |
±2.0%±30mA |
|
|
50/60Hz |
40-100Hz |
|
AC 400mA |
AC 1mV/A |
±2.0%±0.5mA |
±2.5%±0.5mA |
|
50/60Hz |
40-400Hz |
|
AC 4A (0-400mA) |
AC 100mV/A |
±2.5%±8mA |
±3.0%±8mA |
AC 4A (400mA-4A) |
AC 100mV/A |
±2.5%±3mA |
±3.0%±6mA |
AC 30A |
AC 10mV/A |
±2.0%±30mA |
±2.5%±30mA |
1 400mA range is not designed for DC current measurement. The accuracy is
±2.5%±1mA, and it is only used for reference. In the DC400mA range, the output would
show 0.4mV drifting (rdg, rdg+0.1, rdg+0.2, rdg+0.3, rdg+0.4, rdg+0.3, rdg+0.2, rdg+0.1, rdg,...).
Range |
Sensitivity |
DCA 4A |
0.1mV/1mA DC |
DCA 40A |
0.1mV/10mA DC |
ACA 400mA |
0.1mV/0.1mA AC |
ACA 4A |
0.1mV/1mA AC |
ACA 40A |
0.1mV/10mA AC |
General Specifications:
Conductor Size: |
23mm max. (approx.) |
Dimension (Box): |
183 (L) x 61.3 (W) x 35.6 (H) mm 7.2" (L) x 2.5" (W) x 1.4" (H) |
Battery Type: |
two 1.5V SUM-3 |
Range Selection: |
Manual |
Power Consumption: |
10 mA (approx.) |
Low Battery Indication: |
red LED |
Operating Temperature: |
-10°C to 50°C |
Operating Humidity: |
less than 85% relative |
Altitude: |
up to 2000M |
Storage Temperature: |
-20°C to 60°C |
Storage Humidity: |
less than 75% relative |
Weight: |
190g (battery included) |
Accessories: |
Carrying bag x 1 Users manual x 1 Batteries 1.5V x 2 Test leads |